UNIMO Technology UDR-204 Manuel d'instructions Page 301

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 358
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 300
Calibration, Maintenance, Troubleshooting and Warranty
Calibration Setup
Rev. B MultiMode SPM Instruction Manual 281
15.2 Calibration Setup
15.2.1 Check Scanner Parameter Values
If the system's original scanner parameters are deleted, copy the scanner parameters from the
software CD shipped with every system. Individually purchased scanners are shipped with a head/
scanner disk containing backup les, or a hard copy of the scanner parameters. In the event that
les are not found, fax or call Veeco for scanner calibration records.
15.2.2 Align Calibration Reference
Load the silicon calibration reference into the SPM. The reference will need to be aligned with the
microscope scanner so that the tip scans parallel to the reference’s features with the Scan angle set
at 0 degrees.
Align the reference within approximately 2 degrees of perpendicularity to the scan axes.
Scan angle = 0°
Vue de la page 300
1 2 ... 296 297 298 299 300 301 302 303 304 305 306 ... 357 358

Commentaires sur ces manuels

Pas de commentaire